This manual describes how to operate the Planmeca Proline XC Panoramic X-ray unit. Please read these instructions thoroughly before using the unit. Planmeca Proline XC panoramic X-ray fulfills the requirements of Directive 93/42/EEC. NOTE This manual is valid for software revisions: PK 6.54 or later and PG 6.52 or later. Planmeca FIT CAD System User Manual. Download Operation & user's manual of Planmeca PlanScan Dental equipment, Medical Equipment for Free or View it Online on All-Guides.com. This version of Planmeca PlanScan Manual compatible with such list of devices, as: PlanScan, FIT CAM System, FIT CAD System, Emerald S, Emerald. A-2 Planmeca Proline CC panoramic x-ray Technical Manual 2 MANUAL VERSIONS PLANMECA pursues a policy of continuous product development. Although every effort is made to produce up-to-date product documentation this manual should not be regarded as an infallible guide to current production specifications. Planmeca Proline XC Panoramic X-ray User's manual. Page 54: Cleaning PLANMECA service technician once a year or after every 10,000 exposures if this is sooner. Please refer to the Planmeca Proline XC Panoramic X-ray technical manual for complete servicing information. 50 Planmeca Proline XC Panoramic X-ray. Page 55: Disposal Of The Unit.
Technical manual 1.4 User's statement for Planmeca Intra Radiation leakage technique factors The maximum rated peak tube potential is 70 kV and the maximum rated continuous tube current is 0.53 mA for the maximum rated peak tube potential. Minimum filtration The radiation port contains an added 1.0 mm aluminium filtration. The measured half. Planmeca Proline XC Cephalostat 5 PLANMECA PROLINE XC PAN/CEPH X-RAY UNIT User's manual 4.3 Removing and replacing the ear posts and the nasial positioner 4.4 Emergency stop button The emergency stop button is located on the underside of the vertical carriage. Autodesk revit architecture 2011 64bit free download. Pressing the emergency stop button stops all movements and radiation emition. HIGH-QUALITY IMAGES. Planmeca Proline XC features advanced panoramic x-ray technology, including an adjustable focal trough for anatomically correct images, enhanced imaging geometry to eliminate redundant shadows and ghost images, Automatic Cervical Vertebrae (ACV) to minimize the shadow of the cervical vertebrae, and Automatic Gain Control (ACG) for optimized dental radiography quality.
DESCRIPTION
Proline XC X-rays are designed for full-view patient positioning.Proline XC X-rays are designed for full-view patient positioning using a three laser positioning light system and a Graphic User Interface with easy-to-follow, color-coded menu options. The XC's focal trough adjusts to the patient's jaw shape and size providing excellent image quality. In addition, the Proline XC can be upgraded with future software features by simply replacing a software chip! With over 10+ programs from which to choose, the XC is definitely an economical solution for excellent dental imaging.
FORUMSView All (8)
Ask a New Question0Replies10 months ago | 10 months agotroubleshooting message on computer after taking pano I took a pano today and the pano worked but did not come up on the computer and the message says no license on fileReply |
-Del Mar View Dental Care 2 years ago | 2 years agoit wont rotate properly, off center at start up. It wont finish it wont rotate properly, off center at start up. It wont finishReply |
-bwgantt 3 years ago | 3 years agoPlanmeca ProMax We have a Promax that the images produces a white line at mid-line during exposure. We are exporing possible solutions to the white out at mid-line. Reply |
DOCUMENTS / MANUALSView All
VIDEOS
FEATURES
- Open and easy patient access
- Comfortable and stable patient supports
- Side entry and open view for practical and precise patient positioning
- Triple laser beam system for accurate alignment of reference anatomical landmarks
- TheGraphicUserInterface(GUI)for intuitive selection of exposure program and parameters
- With state-of-the-art direct digital imaging the image is available for diagnosis immediately after exposure
Additional Specifications
Generator: Constant potential, microprocessor controlled, operating frequency 80 kHz
X-ray tube: D-052SB
Focal spot size: 0.5 x 0.5 mm
Total filtration: 2.5 mm Al
Anode voltage: 60 - 80 kV
Anode current: 4 - 12 mA DC
Exposure time:
Pan-2.5 - 18 s
Ceph- 0.2 - 5 s
Planmeca 2002 cc proline error
September 02, 2014 10:10
Planmeca Proline Xc User's Manual Guide
Hello, I've the Planmeca Proline 2002 cc Ortho and the machine reports me the Error 57 in the display. I've turn off and then turn on the Ortho, but the error continues.
Can anyone help me?
Thanks!
re: Planmeca 2002 cc proline error
September 03, 2014 09:01
remove the handswitch and turn unit off and on and see if it clears? could be a bad switch closed all the time
Log inor RegisterPlanmeca Proline Xc Calibration
to rate and post a comment
re: Planmeca 2002 cc proline error
September 03, 2014 09:01
Error 57 hand switch activated during switch on check this first
re: re: Planmeca 2002 cc proline error
September 03, 2014 12:54
Is the hand switch the button you press to take a radiography? I don't have this piece instaled. I only have the autoprint machine instaled. Sometimes after the error 57 appears HE 2 it seems like the
Keyboard doesn't respond to the key press. Should I connect the handswitch? How can i install it?
Thank you!!
to rate and post a comment
re: Planmeca 2002 cc proline error
September 03, 2014 09:02
Hi, you have problem in exposure button, praticaly is switched on. Open the exposure botton and also you can have problem in the extention cable. Put out the cable of extention exposure button and the error must deasapeare.
Send me a mail, and i send you technical manual.
Boian
to rate and post a comment
re: re: Planmeca 2002 cc proline error
September 03, 2014 12:54
Is the hand switch the button you press to take a radiography? I don't have this piece instaled. I only have the autoprint machine instaled. Sometimes after the error 57 appears HE 2 it seems like the
Keyboard doesn't respond to the key press. Should I connect the handswitch? How can i install it?
Thank you!!
re: re: re: Planmeca 2002 cc proline error
September 03, 2014 05:34
HE 2 Amalgam container is 100% full (only Metasys
or Dürr).
Replace or empty the container immediately.
The suction system cannot be used before the
container is emptied or replaced. See the
User's manual for details.
send me an email address I will send you the service manual, if your keyboard does not respond and is used to initate the exposure it probably is bad
to rate and post a comment
re: re: re: re: Planmeca 2002 cc proline error
September 04, 2014 10:19
Amalgam in x-ray?!?
Log inor Registerto rate and post a comment
re: re: re: re: re: Planmeca 2002 cc proline error
September 05, 2014 09:38
that came from the service manual
re: re: re: re: Planmeca 2002 cc proline error
Planmeca Usa Support Manuals
March 10, 2017 10:08
Planmeca Promax Technical Manual
Hello Lenny, i need this service manual, can you send it also to me please [email protected]
Thanks so much
re: re: re: re: re: Planmeca 2002 cc proline error
Planmeca Promax Error Codes
March 10, 2017 10:33
OK I will send
Log inor Registerto rate and post a comment
re: re: re: re: re: re: Planmeca 2002 cc proline error
Planmeca Instruction For Uses
March 10, 2017 04:46
Planmeca Prostyle Intra Service Manual
Planmeca Promax 3d Service Manual
Planmeca Promax Manual
Thanks a lot